Research Project

Extravascular Optical Coherence Tomography for Evaluation of Microvascular anastomosis

High resolution and real-time 3D evaluation tool for microvascular anastomosis (or vessel re-connection) based on optical coherence tomography

getImageIn this work, we propose a novel dispersion compensation method that enables real-time compressive sensing (CS) spectral domain optical coherence tomography (SD OCT) image reconstruction. We show that dispersion compensation can be incorporated into CS SD OCT by multiplying the dispersion-correcting terms by the undersampled spectral data before CS reconstruction. High-quality SD OCT imaging with dispersion compensation was demonstrated at a speed in excess of 70 frames per s using 40% of the spectral measurements required by the well-known Shannon/Nyquist theory. The data processing and image display were performed on a conventional workstation having three graphics processing units.

 

getImage (2)

Back to top